Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
7200 | 2025-04-24 | Lab none | Tested | Lab none | I tested this sensor in the lab. The serial interface and temperature seem fine but the CH4 is reading -10.00. I downloaded all the data and erased the internal data file. I will try to get advice from Pro-Oceanus. |
7197 | 2025-04-16 | Mayberry CH4 Probe | Removed | Lab none | Removed from Mayberry due to reading -10.00 again. No problems with the mounting this time. |
7171 | 2025-03-26 | Lab none | Installed | Mayberry CH4 Probe | I reinstalled the miniCH4 on the northeast leg of the scaffolding. |
7160 | 2025-03-13 | Lab none | Tested | Lab none | Tested in the Lab. Seems okay. The CH4 was reading near zero 0.02% and I could make it climb by breathing into the filter. |
7158 | 2025-03-07 | Gilbert Tract CH4 Probe Channel | Removed | Lab none | This one started reading -10% CH4 for no apparent reason. It was removed for testing and replaced by sn 38-574-25 |
7131 | 2025-02-18 | Lab none | Installed | Gilbert Tract CH4 Probe Channel | Installed in the GT channel, co-located with the ADCP, EXO, and channel miniDOT. It has a cable running up the ADCP railing to the enclosure for power/comms. |