Equipment and Calibration History

History for Pro-Oceanus MiniCH4 38-576-25

Rec No.DateLoc FromTitleLoc ToNotes
7200 2025-04-24 Lab none Tested Lab none I tested this sensor in the lab. The serial interface and temperature seem fine but the CH4 is reading -10.00. I downloaded all the data and erased the internal data file. I will try to get advice from Pro-Oceanus.
7197 2025-04-16 Mayberry CH4 Probe Removed Lab none Removed from Mayberry due to reading -10.00 again. No problems with the mounting this time.
7171 2025-03-26 Lab none Installed Mayberry CH4 Probe I reinstalled the miniCH4 on the northeast leg of the scaffolding.
7160 2025-03-13 Lab none Tested Lab none Tested in the Lab. Seems okay. The CH4 was reading near zero 0.02% and I could make it climb by breathing into the filter.
7158 2025-03-07 Gilbert Tract CH4 Probe Channel Removed Lab none This one started reading -10% CH4 for no apparent reason. It was removed for testing and replaced by sn 38-574-25
7131 2025-02-18 Lab none Installed Gilbert Tract CH4 Probe Channel Installed in the GT channel, co-located with the ADCP, EXO, and channel miniDOT. It has a cable running up the ADCP railing to the enclosure for power/comms.