Equipment and Calibration

View History

Equipment Entry

Instrument:Data Logger
Manufacture:Campbell
Model No:CR1000
Serial No:45312
Tag No:
Acquisition Date:2024-10-04
Project:USGS-RR
Location on 2024-11-08:Lab,
Notes: Wiring Panel 38615. Was in a box collecting high frequency flux data and phenocam images., Brought to UCB by Kat Cone for the handoff of Rush Ranch

Documents

No documents found

History

Rec No.DateLoc FromTitleLoc ToNotes
7012 2024-11-08 Lab none Upgrade Lab none I am having trouble getting FTP and ping to work on this logger. I've ruled out problems with the NL115/116, cables and ethernet switch. I could connect to the CR1000 with an IP port. It was running OS v32.02 and the newest OS is v32.07. I updated the OS hoping it would wipe whatever setting might be interfering. Nope - no luck. All the hardware works just fine on another CR1000.