Field Notes
<--2018-02-28 09:20:00 | 2018-03-19 12:25:00-->Phenocam link: Searching...
Sherman Wetland: 2018-03-05 01:37:00 (DOY 64)
Author: Patty Oikawa
Others: Jennie Bahramian
Summary: Calibrated LGR, replaced filters, downloaded data
Arrived 1:37pm First opened box: CO2 418ppm D13C -14 per mil D18O = -70 per mil H2O = 322.6ppm Gas Temp=44.85C Gas Pressure = 119.78 Torr Laser response = 12.85 us Laser adjust looked a little off Initial Laser voltage=2.035 Post adjustment=2.0376 Changed out all filters on tower inlets Connected NOAA standard tank PSI=1800 2nd stage PSI=10 Decided to not “calibrate†laser, best to do that in the lab with temperature controlled conditions First ran NOAA, old standard (from Joe) and new standard (Praxair CO2 400ppm) Set new program, sampling from new standard every 2 hours Sequence: Valve, Sec, ID 2, 300, Ref gas 1, 6600, tower 4, 300, 1.5m 3, 300, 3m Downloaded data |
0 photos found
7 sets found
Can't check MBWPEESWGT_mixed
Can't check MBEEWPSWEPGT_cam
15 SW_met graphs found
4 SW_flux graphs found
Can't check MBSWEEWPBCBA
Can't check SW_Temp
14 SW_Temp graphs found
No data found for SW_arable.