Author: Patty Oikawa
Others:
Summary: Re-established remote connection with LGR, cal gas tank turned off, filters need replacement
I arrived at 1:30pm, it was sunny with a breeze. The bungee cords had blown off of the box as well as part of the tarp. Must have been really windy, need a better way to secure. Cal gas was really low but still showing about 30psi. I turned off tank and took ref gas out of MIU sequence. Need to bring new cal gas next time. Laser alignment wasn’t perfect. D13C from tower is -13 per mil with Laser adjust 2.035 Adjusted laser to 2.040; D13C from tower now -14 per mil (not noticeable difference) Had to re-start the hotspot and teamviewer started working again. I suppose the hotspot does not deal with power outages well but I'm not sure why that is... I downloaded data from 11/27 to 12/28. The tower inlet filter looked dirty but I didn’t have any replacements with me so I swapped with the 3m inlet which was clean. Need to bring more filters next time. D13C from tower after swapping filter was still around -14 per mil--seems the dirty filter wasn't causing much fractionation yet. |