Equipment and Calibration History

History for Kipp & Zonen NR-LITE 990354

Rec No.DateLoc FromTitleLoc ToNotes
5933 2022-08-10 Lab none Installed Experiment Installed on the back of Robert's e-bike for his urban heat island transect in Oakley
4196 2019-10-01 Lab none Calibration Lab none New calibration factor of 13.4 uV/(W/m2).

Tested this sensor at Oxford Tract against AMP 4-band radiometer CNR4 sn 16084. See RT Report "Oxford Tract Radiometer Comparison - 09/2018" for details. https://nature.berkeley.edu/biometlab/bmetdata/rtreport.php?screen=view&id=27
4047 2019-06-18 Lab none Calibration Lab none Tested this sensor at Oxford Tract against AMP 4-band radiometer CNR4 sn 16084. See RT Report "Oxford Tract Radiometer Comparison - 06/2018" for details. I calculated a calibration factor of 12.7 uV/(W/m2), but more testing needs to be done because there was a large calibration difference when testing the sensor upside-down (15.4 uV/(W/m2)).
3776 2018-10-30 Lab none Location Lab none This sensor is on the tram and has probably been there the last 10 years.
2329 2002-06-06 Lab none Installed Tonzi-Floor Installed on CR23x3 with the soil moisture probes at Tonzi.
4049 1999-07-29 Factory Calibration Lab none Initial factory calibration 12.9 uV/(W/m2) 1999-07-29