Equipment and Calibration History

History for Environmental Sensors Inc. PR K Type 8 00000FED8688

Rec No.DateLoc FromTitleLoc ToNotes
199 2010-09-03 Factory Disposed Disposed Report from vendor not repairable:
s/n 00000FED8688
Condition on arrival: cracks in the epoxy of top segment (segment 1), and signs of gaps between rails (segment 1 and segment 3). RF Ground and Ctrl Ground pins open circuit, only diode 2 control line passes resistance checks. No diode waveform from +1 diode board, so only segments 3 and 4 readings seem to be okay.
175 2010-07-07 Lab none Repair Factory Sent to ESI - Environmental Sensors Inc (www.esica.com)for evaluation and possible repair.
148 2010-06-22 Sherman Island Removed Lab none Removed from the Sherman Island cow pasture. This probe was outside the cow fence and giving bad readings.