Equipment and Calibration History

History for Vaisala GMP221 Y4110008

Rec No.DateLoc FromTitleLoc ToNotes
590 2012-05-02 Factory Disposed Disposed Filter was cracked, probe completely replace - new probe is serial number H1310008
573 2012-03-28 Lab none Shipped Factory Sent to Vaisala for repair. Reporting errors 10, 11 and 12 readings unreasonable.
1290 2010-04-06 Lab none Installed Sherman Island Installed at Sherman Island in Set A at -2cm.
1363 2010-02-11 Lab none Calibration Lab none Trueppm = Probeppm*1.019 + -108.4; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 383ppm, 597ppm, and 2238ppm. Ta=21.8C, P=102.3kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Jaclyn.
14 2009-12-01 Factory Repair Factory Vaisala factory repair and calibration 0-3% - certificate on file.
Problem: Got probe error 10 and 11. No IR source.
Action: Replaced probe circuit board and adjusted unit.
20 2005-05-16 Tonzi Removed Lab none Removed from Tonzi - location at Tonzi unknown