Date: |
2010-04-07 |
From: |
Factory - |
To: |
Lab - none |
Title: |
Repair |
Notes: |
Returned from LGR. LGR bill says pressure controller, filter, and check valve were replaced. Mirrors were cleaned. Running in slow flow mode analyzer is reporting 156.20 Torr and 18.587 usec ring time. |
Entered by: |
|
|
|