Date: |
2013-03-18 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Calibration |
Notes: |
Last calibration: 2012-08-23
Cell Pressure=157.32Torr, Cell Temperature=24.40C, Ring Time=12.694usec
2nd CH4 in Air at psi
CH4 Span=3.875ppm
CH4 = 3.888 -> 3.872 Offset: -13ppb
Notes: Done in the lab after cleaning the 10um filter infront of the pressure controller
|
Entered by: |
|
|
|