Equipment and Calibration History

View Equipment |

View history entry 946 for Los Gatos Research FMA RMT-200 sn: 05-0010

Date: 2013-03-18
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2012-08-23 Cell Pressure=157.32Torr, Cell Temperature=24.40C, Ring Time=12.694usec 2nd CH4 in Air at psi CH4 Span=3.875ppm CH4 = 3.888 -> 3.872 Offset: -13ppb Notes: Done in the lab after cleaning the 10um filter infront of the pressure controller
Entered by: