Date: |
2013-02-21 |
From: |
Twitchell Rice - Gas Analyzer CH4 |
To: |
Twitchell Rice - Gas Analyzer CH4 |
Title: |
Calibration |
Notes: |
Calibration Location: Twitchell
Last calibration: 2012-09-06
RSSI=65%, Optical RH=6%
Ambient P=102.1kPa, Tair=7.87C
Zero Air at 1800psi, 2nd CH4 in Air 4.982 at 1300
CH4 Zero Offset: 0.005ppm
CH4 Span Offset: 0.018ppm
Notes:
|
Entered by: |
|
|
|