Equipment and Calibration History

History for YSI EXO Turbidity 22C100485

Rec No.DateLoc FromTitleLoc ToNotes
6192 2023-02-15 Lab none Returned Returned none Returned to Adina during our Elkhorn Slough visit
6196 2023-01-25 Gilbert Tract none Removed Lab none USGS removed this EXO from the Gilbert Tract channel and stashed it on the bank. Joe retrieved it from the site at the end of the day and brought it back to the lab.
6059 2022-11-01 Lab none Calibration Lab none Turbidity (FNU), Last cal: Unknown
EXO sn 22C100485
EXO/Sensor Firmware: 1.0.86/3.0.4
A factory reset was done first before this calibration was done in the lab.
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 FNU,0.24 FNU,0.00 FNU,0.00 FNU,20.384 °C
Pt2,124.00 FNU,112.11 FNU,123.99 FNU,0.00 FNU,20.389 °C
Pt3,1010.00 FNU,939.77 FNU,1009.99 FNU,0.00 FNU,20.565 °C