Equipment and Calibration History
History for YSI EXO Turbidity 22C100485
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
6192 | 2023-02-15 | Lab none | Returned | Returned none | Returned to Adina during our Elkhorn Slough visit |
6196 | 2023-01-25 | Gilbert Tract none | Removed | Lab none | USGS removed this EXO from the Gilbert Tract channel and stashed it on the bank. Joe retrieved it from the site at the end of the day and brought it back to the lab. |
6059 | 2022-11-01 | Lab none | Calibration | Lab none | Turbidity (FNU), Last cal: Unknown EXO sn 22C100485 EXO/Sensor Firmware: 1.0.86/3.0.4 A factory reset was done first before this calibration was done in the lab. ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 FNU,0.24 FNU,0.00 FNU,0.00 FNU,20.384 °C Pt2,124.00 FNU,112.11 FNU,123.99 FNU,0.00 FNU,20.389 °C Pt3,1010.00 FNU,939.77 FNU,1009.99 FNU,0.00 FNU,20.565 °C |