Equipment and Calibration

View History

Equipment Entry

Instrument:Gas Analyzer
Manufacture:Licor
Model No:LI-7500A
Serial No:75H-2418
Tag No:
Acquisition Date:2022-08-16
Project:Ameriflux
Location on 2024-01-23:Returned,
Notes: Received from Sebastian at Ameriflux for emergency replacement of head that died at GT. Came with box 1039, box brackets, head mount, IRGA cable and temp sensor.

Documents

Manual: LI-7500A_Manual_V1_200912.pdf
Manual: LI-7500A_Manual_v2_201104.pdf
Manual: LI-7500A_Manual_V4_201303.pdf
Manual: LI-7500A_Manual_V5_201310.pdf

History

Rec No.DateLoc FromTitleLoc ToNotes
6541 2024-01-23 Lab none Returned Returned none Returned to Stephen Chan at LBL. Returned with AMP LI-7550 sn AIU-1039, a 7550 thermistor, IRGA cable, and IRGA mount.
6458 2023-11-02 Disposed none Borrowed Lab none Borrowed from AMP for while our LI-7550 sn AIU-1866 is back at Licor for repairs. Came with AMP LI-7550 sn AIU-1039, a 7550 thermistor, IRGA cable, and IRGA mount.
6300 2023-06-13 Lab none Returned Disposed none Returned to Andre Santos from Ameriflux. Returned with box AIU-1039, box brackets, head mount, IRGA cable and temp sensor.
6175 2023-01-19 Gilbert Tract Gas Analyzer CO2/H2O Removed Lab none Removed from Gilbert Tract for calibration. Replaced by sn 75H-2667.
5945 2022-08-18 Lab none Installed Gilbert Tract Gas Analyzer CO2/H2O Installed on the GT tower to replace the one that failed a few days ago and was removed.
5941 2022-08-17 Lab Calibration Lab Last calibration: 2020-03-23
AGC=100.5 SDM=0 DAC1=- DAC2=-
Ambient P=100.2kPa, Tair=24.2C, Tdew=13.6C
Zero Air at 300psi, AMP CA04119 524.77 at 1400 Tdewspan=20.0
CO2 zero -> 0.9786 Offset: -0.7ppm
CO2 span 1.0091 -> 1.0054 Offset: 2.63ppm
H2O zero 0.8245 -> 0.8242 Offset: 0.9ppm
H2O span 1.0166 -> 1.0292 Offset: -0.28ppm
Notes: Borrowed from Ameriflux. Previous cal values retrieved from instrument. Seems the 7500s don't drift much and that our cal methods match whatever methods were used last time.