Equipment and Calibration History

View Equipment |

View history entry 9 for Vaisala GMP222 sn: Z2720015

Date: 2009-12-01
From: Factory -
To: Factory -
Title: Repair
Notes: Vaisala factory repair and calibration 0-1% - certificate on file. Problem: Got probe error 05 and 10. Memory error and no IR source. Action: Replaced probe circuit board and adjusted unit