Equipment and Calibration History
History for Vaisala HMP155 S3420870
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
5239 | 2021-04-15 | Lab none | Installed | Bouldin Alfalfa RH & Temp | Installed on the BA tower after reconnecting sensor to cable screw terminals in the lab. The collar has a metal set screw. |
5235 | 2021-04-13 | Lab none | Repair | Lab none | Reconnected loose wires on cable to connector. The connector has tiny screw terminals, so it was pretty convenient; no soldering needed. Ran for 10 minutes in the lab and values were reasonable. |
5232 | 2021-03-30 | Bouldin Alfalfa RH & Temp | Removed | Lab none | Removed from BA and brought back to the lab for repair. This sensor had fallen to the ground. The cable head needs rewiring. |
5215 | 2021-03-16 | Lab none | Installed | Bouldin Alfalfa RH & Temp | Installed at BA to replace sn G4540021, which was brought back to the lab for testing. |