Date: |
2012-08-16 |
From: |
West Pond - DO Sensor Upper |
To: |
West Pond - DO Sensor Upper |
Title: |
Serviced |
Notes: |
DO-05 was serviced in the field due to continued unexpectedly low readings. The inside was rinsed with vinegar and scrubbed with a tooth brush. The electrolyte was replaced. |