History for DFRobot GL-136 GL-136-01

Rec No.DateLoc FromTitleLoc ToNotes
4397 2020-02-25 Lab Tested Lab I tested this sensor in the lab and could not get it to work. I used the current-to-voltage board provided with the sensor connected to a CR1000. I also connected a multimeter in place of the current-to-voltage board to measure the current. Both setups showed no response even when dunking the sensor in about 1.5m pipe filled with water. I recommend disposal.
4389 2020-02-20 Mayberry 0 Removed Lab none Removed from MB and brought back to the lab b/c it hasn't been comparing well to the regular CSI water level sensor.
4095 2019-09-04 Lab none Installed Mayberry 0 This new sensor is in the water next to the northeast corner post of the scaffolding. It was in about 21cm of water and about 5 cm above the bottom. Test to see if a cheap sensor can do as well as the expensive ones.