Equipment and Calibration

View History

Equipment Entry

Instrument:Pump
Manufacture:KNF Neuberger
Model No:MPU 2096-N86-11.07
Serial No:1/1062115
Tag No:
Acquisition Date:2008-02-01
Project:DOE-TCP
Location on 2010-02-17:Lab,
Notes:

Documents

No documents found

History

Rec No.DateLoc FromTitleLoc ToNotes
65 2010-02-17 Lab none Maintenance Lab none I took this pump apart today to try and clean. After removing bolts from the back of the motor it was clear that it was full of carbon. It took a lot of work to get the pump portion off the front of the motor and was probablly unnecessary. Even with the pump removed, it is difficult and possibly irreversable to open the motor. Instead I blew compressed air through the bolt holes. This seems to clear the insides and the pump once reassembled seemed to run well.
64 2010-02-10 Tonzi-Tower Removed Lab none The CO2 profile system had been shut down for about 10 days to conserve power. On restart this pump began to work but quickly bogged down and stopped. Returned to lab.