Equipment and Calibration History

View Equipment |

View history entry 6900 for Licor LI-7700 sn: TG1-0203

Date: 2024-10-07
From: Lab -
To: Lab -
Title: Calibration
Notes: Calibration Location: Lab Last calibration: 2016-09-09 RSSI=79%, Optical RH=11% Ambient P=99.58kPa, Tair=33.8C Zero Air at 1300psi, AMP FB04035 1.9156 at 850 CH4 Zero Offset: -0.087ppm CH4 Span Offset: 0.0034ppm Notes: First check of instrument from USGS. Last zero was 2016-01-08. Clock 20min fast, set heater and wash to Biomet Default. Head humidity at 38%. Swapped head desiccant and flushed head space to 11%. Initial span used ppb instead of ppm for off by a factor of 1000. Redid span on 10/14 with different tank FB04319, 2.6969ppm CH4, 1000psi - seemed fine.
Entered by: