Equipment and Calibration History

View Equipment |

View history entry 6790 for YSI EXO Wiped Cond/Temp sn: 23D104030

Date: 2024-07-21
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Sp Cond (uS/cm),Last cal: 2024-04-08 EXO sn 22C103416 EXO/Sensor Firmware 1.0.86/3.0.5 CellConstant 0.48 ,Standard,PreCal,PostCal,RawValue,Temp Pt1,5000.0 µS/cm,4976.0 µS/cm,5000.1 µS/cm, 0.0 µS/cm,19.884 °C Note: Used USGS standard from Fondriest that expired Aug 2023. It was stored in fridge w/ cap taped and I brought it close to room temperature before calibration.
Entered by: Daphne Szutu