Equipment and Calibration History

View Equipment |

View history entry 6788 for YSI EXO pH-ORP sn: 22D102641

Date: 2024-07-21
From: Lab - none
To: Lab - none
Title: Calibration
Notes: pH, Last cal: 2024-04-08 EXO sn 22C103416 EXO/Sensor Firmware: 1.0.86/3.0.0 ,Standard,PreCal,PostCal,RawValue,Temp Pt1,7.01 pH,7.01 pH,7.01 pH,-10.1 mV,20.611 °C Pt2,10.06 pH,10.29 pH,10.06 pH,-189.7 mV,20.732 °C Pt3,4.00 pH,3.95 pH,4.03 pH,157.7 mV,20.771 °C 4-7Slope,4-7mVPerDecade,7-10Slope,7-10mVPerDecade 167.8 mV,55.74 mV,179.6mV,58.87 Note: Fresh pH/ORP module. Used opened standards from Dec 2023.
Entered by: Daphne Szutu