Date: |
2024-05-28 |
From: |
Factory - none |
To: |
Factory - none |
Title: |
Calibration |
Notes: |
Device failed in calibration due to faulty sensor.
Action(s) taken
Sensor module, body of the GMP343 and filter changed.
Operation tested, adjustment made and calibrated.
Calibration certificate H30-24220020 issued. |
Entered by: |
Joe Verfaillie |
|
|