Date: |
2024-04-08 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Calibration |
Notes: |
ORP (mV), Last Cal: 2024-01-22
EXO sn 22C103416
EXO/Sensor Firmware 1.0.86/3.0.0
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,226.7 mV,228.5 mV,226.7 mV,0.0 mV,23.000 °C
Note: Used Zobell solution (mixed 2022-09) in small brown plastic bottle. Waited to calibrate until change <0.1mV/10sec (a few min past the point when "Unstable" changes to "Stable"). |
Entered by: |
Daphne Szutu |
|
|