Date: |
2024-04-08 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Calibration |
Notes: |
pH, Last cal: 2024-01-22
EXO sn 22C103416
EXO/Sensor Firmware: 1.0.86/3.0.0
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,7.01 pH,7.06 pH,7.01 pH,-8.2 mV,22.000 °C
Pt2,10.06 pH,10.15 pH,10.08 pH,-179.0 mV,22.000 °C
Pt3,4.00 pH,4.14 pH,4.02 pH,154.6 mV,22.000 °C
4-7Slope,4-7mVPerDecade,7-10Slope,7-10mVPerDecade
162.8 mV,54.09 mV,170.7mV,55.98
Note: Waited for each point to change <0.1mV/10sec for calibration (a few minutes past point where "Unstable" changes to "Stable"). Used opened standards from May 2023.
|
Entered by: |
Daphne Szutu |
|
|