Date: |
2024-01-22 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Calibration |
Notes: |
ORP (mV), Last Cal: 2023-10-16
EXO sn 22C103416
EXO/Sensor Firmware 1.0.86/3.0.0
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,230.4 mV,227.6 mV,223.3 mV,0.0 mV,21.275 °C
Note: Poured Zobell solution (mixed 2022-09) into small brown plastic bottle for this calibration. |
Entered by: |
Daphne Szutu |
|
|