View history entry 6550 for YSI EXO Wiped Cond/Temp sn: 23D104030

Date: 2024-01-22
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Sp Cond (uS/cm),Last cal: 2023-08-17 EXO sn 22C103416 EXO/Sensor Firmware 1.0.86/3.0.5 CellConstant 0.46 ,Standard,PreCal,PostCal,RawValue,Temp Pt1,8420.0 µS/cm,8507.6 µS/cm,8420.0 µS/cm, 0.0 µS/cm,20.226 °C Note: Used 10,000 uS/cm standard from fridge (brought it to room temperature first). Measured temp-compensated conductivity with handheld YSI probe
Entered by: Daphne Szutu