Equipment and Calibration History
History for Decagon-METER SRS-Ni NDVI 2046503354
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
6415 | 2023-09-18 | Lab none | Loaned | Loaned none | Loaned to Gracie Pearsall at UCSC for a informational display at Elkhorn Reserve |
6180 | 2023-01-25 | Lab none | Tested | Disposed none | Removed the spliced cable and tested in the lab. Still no response on the terminal emulator. This sensor is probably dead. It's weird that both incoming and outgoing sensors went bad at the the same time. Both sensors had spliced cables and were connected to the same C port and the same pluggable power block. One of the spliced wires for data (red) had a few cracks in the plastic sheath; maybe it shorted against the bare G wire and killed both sensors. |
6171 | 2023-01-18 | Mayberry Narrow Band NDVI Reflected | Removed | Lab none | Removed from MB for testing in the lab. It was not responding on the terminal emulator. |
2920 | 2017-06-05 | Lab none | Installed | Mayberry Narrow Band NDVI Reflected | Installed at outgoing position on Mayberry Tower. |
2986 | 2017-06-04 | Lab none | Maintenance | Lab none | The SDI-12 address was set to E indicating a hemispherical incoming orientation and the head was labeled with sn, address, and 'out'. |