Date: |
2022-11-30 |
From: |
Hill Slough - |
To: |
Hill Slough - |
Title: |
Calibration |
Notes: |
Calibration Location: Hill Slough
Last calibration: 2022-06-08
RSSI=71%, Optical RH=26%
Ambient P=101.4kPa, Tair=10.3C
Zero Air at 1650psi, AMP CA04396 1.9317 at 1100
CH4 Zero Offset: .005ppm
CH4 Span Offset: .0003ppm
Notes: After calibration we flushed headspace with zero air until optics RH = 8%. We swapped in fresh drierite. The current thermocouple assembly is missing an o-ring (has putty instead in the o-ring groove), so we swapped in a brand-new thermocouple assembly.
|