Date: |
2022-07-12 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Calibration |
Notes: |
pH, Last cal: Unknown
EXO sn 22C103416
EXO/Sensor Firmware: 1.0.86/3.0.0
Slope is lower than what it should be for a new sensor,Standard,PreCal,PostCal,RawValue,Temp
Pt1,7.00 pH,7.02 pH,7.00 pH,-1.4 mV,20.000 °C
Pt2,10.00 pH,9.89 pH,10.00 pH,-167.8 mV,20.000 °C
Pt3,4.00 pH,4.39 pH,4.02 pH,151.1 mV,20.000 °C |
Entered by: |
Joe Verfaillie |
|
|