Date: |
2021-12-01 |
From: |
Hill Slough - |
To: |
Hill Slough - |
Title: |
Calibration |
Notes: |
Calibration Location: Hill Slough
Last calibration: 2021-06-01
RSSI=72%, Optical RH=38%
Ambient P=101.6kPa, Tair=16.6C
Zero Air at 950psi, AMP CA04396 1.9317 at 1500
CH4 Zero Offset: -.066ppm
CH4 Span Offset: .0633ppm
Notes: Relatively still day. After calibration, flushed out headspace until optics RH=10% and swapped out desiccant bottle for one with fresh drierite. Zero air tank pressure uncertain--wonky regulator.
|