View history entry 5539 for Texas Electronics TR-5251 sn: 40415-1006

Date: 2021-10-21
From: Mayberry - Precipitation
To: Mayberry - Precipitation
Title: Tested
Notes: We dripped 240ml of water through the bucket using a peristaltic pump powered with a lithium battery. This should result in 50 tips in about 20 minutes. We got 54 tips and adjusted its set screws 1.5 turns CCW (1/2 turn of both screws is a 2-3% adjustment).
Entered by: Daphne Szutu