Date: |
2021-06-01 |
From: |
Lab - |
To: |
Lab - |
Title: |
Calibration |
Notes: |
Last calibration: 2021-03-03
AGC=100 SDM=0 DAC1=- DAC2=-
Ambient P=100.5kPa, Tair=21.8C, Tdew=10.7C
Zero Air at 300psi, AMP CA04119 524.77 at 1800 Tdewspan=16.0
CO2 zero 0.9003 -> 0.9004 Offset: 0.6ppm
CO2 span 0.9978 -> 1.0016 Offset: -2.67ppm
H2O zero 0.8461 -> 0.8444 Offset: -9.7ppm
H2O span 0.9906 -> 0.9808 Offset: 0.2ppm
Notes: I reset the SS from 95 to 100 before calibration.
|
Entered by: |
|
|
|