Date: |
2020-11-17 |
From: |
Lab - |
To: |
Lab - |
Title: |
Calibration |
Notes: |
Calibration Location: Lab
Last calibration: 2020-05-22
RSSI=72%, Optical RH=22%
Ambient P=100.2kPa, Tair=20.5C
Zero Air at 575psi, AMP CA4396 1.9497 at 400
CH4 Zero Offset: -.064ppm
CH4 Span Offset: .011ppm
Notes: Swapped optics drierite after calibration and flushed headspace with zero air. New optics RH was 11%.
|
Entered by: |
|
|
|