Equipment and Calibration
View History
Equipment Entry
|
Documents
Calibration: Hukseflux_NR01_2466.pdfPackingSlip: ps_Campbell_20150928.pdf
Report: SW 4-way comparison - 2021 - 2021-02-25
History
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
5503 | 2021-09-02 | Sherman Wetland Net Radiometer 4-way | Moved | Gilbert Tract Net Radiometer 4-way | Moved from SW to GT with the decomissioning of SW and the beginning of GT |
2291 | 2016-05-24 | Lab none | Installed | Sherman Wetland Net Radiometer 4-way | Installed on the Sherman Wetland tower at the initial set up of the site. |
1972 | 2015-04-10 | Factory | Calibration | Factory | Initial factory calibration: 4 components Component SWin sn4057 2015-03-26 a0 = 0 a1 = 13.92 uV/(W/m2) Component SWout sn4058 2015-03-26 a0 = 0 a1 = 15.09 uV/(W/m2) Component LWin sn3998 2015-02-03 a0 = 0 a1 = 13.54 uV/(W/m2) Component LWout sn3999 2015-02-03 a0 = 0 a1 = 14.14 uV/(W/m2) |