Equipment and Calibration History

View Equipment |

View history entry 4485 for Licor LI-7700 sn: TG1-0211

Date: 2020-05-21
From: Sherman Wetland-Temp1 - Gas Analyzer CH4
To: Sherman Wetland-Temp1 - Gas Analyzer CH4
Title: Calibration
Notes: Calibration Location: Sherman Wetland Temp Last calibration: 2019-10-22 RSSI=73%, Optical RH=23% Ambient P=101.3kPa, Tair=25.8C Zero Air at 700psi, AMP CA3923 3.121 at 700 CH4 Zero Offset: 0.023ppm CH4 Span Offset: -0.005ppm Notes: Breezy day. Swapped optics drierite after calibration and used cal gas to dry head space. New optics RH was 8%.
Entered by: