View history entry 4393 for Licor LI-7500A sn: 0418

Date: 2020-02-19
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2019-11-18 AGC=93.2 SDM= DAC1= DAC2= Ambient P=100.8kPa, Tair=28.1C, Tdew=8.2C Zero Air at 800psi, AMP CA4396 408.97 at 550 Tdewspan=16.00 CO2 zero 0.8991 -> .8994 Offset: 1.12ppm CO2 span 0.9988 -> .9987 Offset: -.74ppm H2O zero 0.8482 -> .8456 Offset: -5.4ppm H2O span 0.9812 -> .9845 Offset: -.06ppm Notes: