Equipment and Calibration History

View Equipment |

View history entry 421 for Los Gatos Research FMA RMT-200 sn: 05-0010

Date: 2011-10-10
From: Lab - none
To: Lab - none
Title: Maintenance
Notes: No flow in low flow mode. The internal 2um after the pressure controller seemed okay. I disassembled the inlet tubing and fittings to get a 10um filter out from infront of the pressure controller. This was completely plugged. I rinsed it with degreaser and DI water and dried it by blowing air through it. Flow was better but still seemed low. I reinstalled the filter and reassembled the inlet tubing. Now is slow flow the internal pressure is 156Torr.
Entered by: