Equipment and Calibration
View History
Equipment Entry
|
Documents
Calibration: Forerunner_GP001_20140026.pdfReport: Eosense CO2 Probe Compare - 2018-09-07
Report: eosGP comparison - 11/2023 - 2023-11-16
History
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
5236 | 2021-04-13 | Lab none | Tested | Lab none | Tested in the lab and the CO2 signal responded to my breath. I'll keep it running for a while and see. |
5109 | 2021-02-04 | West Pond CO2 Probe B | Removed | Lab none | Removed from WP and brought back to the lab b/c this sensor has been reading noise since late Nov |
4107 | 2019-09-18 | Lab none | Installed | West Pond CO2 Probe B | Reinstalled at the northwest corner of the scaffolding. The water level is low so it is sitting in a puddle for the time being. |
4105 | 2019-09-16 | Lab none | Tested | Lab none | Tested against 2 good sensors (1% and 2%) in the lab test chamber with atmospheric CO2 and with an exhalation. Seemed to respond similarly to the other sensors. This sensor has degraded threads, so you can't get the cable off. I will seal up the connection again. I'm not sure why it was maxing out in the field, but seems fine now. Maybe some condensation inside evaporated? |
4048 | 2019-08-19 | West Pond CO2 Probe B | Removed | Lab none | Removed from WP and brought back to the lab for testing out (it was maxing out at 2% even in the air). |
3791 | 2018-11-01 | Lab none | Installed | West Pond CO2 Probe B | Installed at WP to replace sn GP20140018, which was brought back to the lab for too-high readings. |
3354 | 2018-03-07 | Twitchell Alfalfa | Removed | Lab none | Removed from the -8cm location at Twitchell Alfalfa because the readings have been funky for a while. Probably will be replaced at a later date. |
1935 | 2015-08-19 | Lab none | Installed | Twitchell Alfalfa | Installed at the 8cm level at Twitchell Alfalfa to replace sn GP2015ex03 that was dropping out. |
1897 | 2015-06-25 | Mayberry CO2 Probe B | Removed | Lab none | Removed from the middle location at Mayberry due to corroding connector. Replaced by sn GP20140025 |
1726 | 2015-02-05 | Lab none | Installed | Mayberry CO2 Probe B | Installed in the middle position of the stringer at Mayberry. This sensor has copper mesh over the filter and replaces sn GO20140024 |