Equipment and Calibration History

View Equipment |

View history entry 3929 for Licor LI-7500A sn: 0418

Date: 2019-04-22
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2019-01-07 AGC=95 SDM= DAC1= DAC2= Ambient P=100.6kPa, Tair=24.2C, Tdew=11.4C Zero Air at 1100psi, AMP Air 408.97 at 700 Tdewspan=16.00 CO2 zero 0.8992 -> .8992 Offset: -.04ppm CO2 span 1.0033 -> 1.0011 Offset: 1.21ppm H2O zero 0.8465 -> .8464 Offset: -.4ppm H2O span 0.9832 -> .9866 Offset: -.07ppm Notes:
Entered by: