Equipment and Calibration History

View Equipment |

View history entry 3794 for Forerunner-Eosense GP-001 sn: GP20140018

Date: 2018-11-06
From: Lab - none
To: Lab - none
Title: Tested
Notes: Connected to a CR6 and recorded 10s data. The sensor was maxing out (reading ~5V) on both 0-2% and 0-6% ranges. The connection between sensor and cable seemed secure and clean. The sensor should go back to the factory.
Entered by: Daphne Szutu