Equipment and Calibration History

View Equipment |

View history entry 3731 for Licor LI-7500A sn: 0418

Date: 2018-10-10
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2018-04-13 AGC=93.2 SDM=0 DAC1=0 DAC2=0 Ambient P=99.98kPa, Tair=29.11C, Tdew=15.00C Zero Air at 1500psi, AMP Air 408.97 at 1200 Tdewspan=16.00 CO2 zero 0.9001 -> .8982 Offset: -7.43ppm CO2 span 1.0057 -> .9965 Offset: 4.86ppm H2O zero 0.8426 -> .8454 Offset: 16.8ppm H2O span 0.9892 -> .9813 Offset: 0.16ppm Notes: calibrated by Katrina
Entered by: