Equipment and Calibration History
History for Kipp & Zonen NR-LITE 990354
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
6899 | 2024-10-14 | Lab none | Removed | Lab none | Recovered from the bike package. This sensor went to Burning Man in 2023 and has been sitting idle in Robert's office ever since. Stored in box of NR-Lites next to Daphne's desk. |
5933 | 2022-08-10 | Lab none | Installed | Experiment | Installed on the back of Robert's e-bike for his urban heat island transect in Oakley |
4196 | 2019-10-01 | Lab none | Calibration | Lab none | New calibration factor of 13.4 uV/(W/m2).
Tested this sensor at Oxford Tract against AMP 4-band radiometer CNR4 sn 16084. See RT Report "Oxford Tract Radiometer Comparison - 09/2018" for details. https://nature.berkeley.edu/biometlab/bmetdata/rtreport.php?screen=view&id=27 |
4047 | 2019-06-18 | Lab none | Calibration | Lab none | Tested this sensor at Oxford Tract against AMP 4-band radiometer CNR4 sn 16084. See RT Report "Oxford Tract Radiometer Comparison - 06/2018" for details. I calculated a calibration factor of 12.7 uV/(W/m2), but more testing needs to be done because there was a large calibration difference when testing the sensor upside-down (15.4 uV/(W/m2)). |
3776 | 2018-10-30 | Lab none | Location | Lab none | This sensor is on the tram and has probably been there the last 10 years. |
2329 | 2002-06-06 | Lab none | Installed | Tonzi-Floor | Installed on CR23x3 with the soil moisture probes at Tonzi. |
4049 | 1999-07-29 | Factory | Calibration | Lab none | Initial factory calibration 12.9 uV/(W/m2) 1999-07-29 |