Date: |
2017-05-31 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Calibration |
Notes: |
Calibration Location: Lab
Last calibration: 2016-10-11
RSSI=71.2%, Optical RH=14%
Ambient P=100.3kPa, Tair=C
Zero Air at 1500psi, NOAA Air 1.89919 at 1600
CH4 Zero Offset: -0.1ppm
CH4 Span Offset: -0.002ppm
Notes: Replaced temperature component. Replaced desiccant and flushed cavity with zero air. Optics RH was 26% before replacement and 14% afterwards.
|
Entered by: |
|
|
|