Date: |
2017-02-24 |
From: |
Factory - |
To: |
Lab - none |
Title: |
Returned |
Notes: |
Returned from Campbell after repair for: 12V screw terminal damage, internal thermistor damage.
Campbell says: Able to verify customer problem. Replaced terminal connector (P22C) 12V connector. Replaced internal thermistor (Z16) due to leads being broken off. Bench tested unit, calibrated unit with documentation, final tested to verify correct operation. |
Entered by: |
Joe Verfaillie |
|
|