Date: |
2016-08-15 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Calibration |
Notes: |
Last calibration: 2016-02-22
AGC=49 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.59kPa, Tair=22.61C, Tdew=12.50C
Zero Air at 500psi, God CO2 421.22 at 500 Tdewspan=16.00
CO2 zero 0.9013 -> 0.9012 Offset: -0.60ppm
CO2 span 1.0076 -> 1.0070 Offset: 0.38ppm
H2O zero 0.8564 -> 0.8558 Offset: -3.6ppm
H2O span 1.007 -> 0.9911 Offset: 0.34ppm
Notes: At first the diagnostic value was 24 with flags set on PLL, detector and chopper. Reconnecting the head fixed the problem.
|