Date: |
2015-03-03 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Calibration |
Notes: |
Patty calibrated the LI6400, zero and span.
Zero
CO2: -22uml (after calibration: 0.0uml )
H2O: -1.04mml (after calibration: 0.003mml)
Span
Gas = 421.11ppm
CO2: 427.8uml (after calibration: 421.0uml)
Dew point gen=15.95 degrees
H2O: 16.27 (after calibration: 15.95) |
Entered by: |
Patty Oikawa |
|
|