Equipment and Calibration History

View Equipment |

View history entry 1544 for Licor LI-7700 sn: TG1-0211

Date: 2014-09-10
From: Twitchell Rice - Gas Analyzer CH4
To: Twitchell Rice - Gas Analyzer CH4
Title: Calibration
Notes: Calibration Location: Twitchell Last calibration: 2014-03-07 RSSI=69%, Optical RH=22% Ambient P=kPa, Tair=25.9C Zero Air at 1400psi, 2nd CH4 4.995 at 800 CH4 Zero Offset: 0.07ppm CH4 Span Offset: -0.075ppm Notes: Very warm day with no wind. Offsets seem a little large.
Entered by:
Field notes