View history entry 1494 for Los Gatos Research FMA RMT-200 sn: 05-0010

Date: 2011-12-18
From: Twitchell Rice -
To: Twitchell Rice -
Title: Calibration
Notes: All was running. The LGR read: 9:04PST, 31.36C, 97.41Torr, 13.973usec, 10Hz, 2.964ppm I calibrated the LGR with secondary tank with 3.276ppm CH4 at 250psi. Last cal was 2011-11-22 Pre cal: 9:12PST, 30.18C, 124.54Torr, 13.972usec, 1Hz, 3.200ppm Pressure was a little low but okay. After cal the LGR read 3.271ppm. I adjusted the bypass valve to bring the pressure to 146Torr in high flow mode. I returned the LGR to 10Hz before leaving.