View history entry 1489 for Los Gatos Research FMA RMT-200 sn: 05-0010

Date: 2010-05-13
From: Twitchell Rice -
To: Twitchell Rice -
Title: Calibration
Notes: The LGR was reporting: 8:27PDT, 22.89C, 120.58Torr, 16.781usec, 10Hz, 1.858ppm I adjusted the bypass valve to correct the pressure. I calibrated the LGR with 3.276ppm tank with 700psi. Last cal was March 22, 2011 Precal: 8:47PDT. 24.11C, 156.50Torr, 16.776usec, 1Hz, 3.290ppm After cal 3.276ppm Before I left the LGR read: 8:52PDT, 24.49C, 150.77Torr, 16.774usec, 10Hz, 1.883ppm
Entered by: Sara Knox