Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
2323 | 2015-09-02 | Mayberry DO Sensor C | Removed | Lab none | |
792 | 2011-03-01 | Lab none | Installed | Mayberry DO Sensor C | DO-03 was installed in the Tules at the new Mayberry Tule Tower. |
276 | 2011-01-24 | Lab none | Calibration | Lab none | On January 24, 2011, Jaclyn Hatala calibrated 2 CS511 Sensorex dissolved oxygen sensors in the lab.
1. I filled the caps with water, to make sure that the membrane was secure and not leaking. 2. I emptied the caps and filled them with the electrolyte solution provided. 3. I attached the two sensors to a datalogger with a multiplier of 1.0, and recorded data every 10 seconds for 15 minutes with the sensors out of water 4. I calculated the multiplier by the equation: M=P/R provided in the manual, where P is the given atmospheric O2 concentration given the room temperature and pressure (in this case for 24C and sea level, 8.09ppm) 5. The sensor was labeled DO_3: coef 02352ppm/mV calculations are in DOcal_24Jan2011.xlsx. |